The Labindia UV-3000 series of UV-Visible spectrophotometers is built using state-of-the-art technology, representing a new generation of double beam spectrophotometer designed for accurate, high-resolution absorbance measurement.
The instrument delivers precise UV-Vis absorbance readings with advanced optical performance, covering a broad wavelength range suited to research, pharmaceutical, and quality control applications.
Key Applications
- Analysis of foods, drugs, and agricultural products.
- Medical care, public health, and environmental protection.
- Life sciences research.
- Photometric measurement, spectrum scans, kinetic measurements, quantitative determination, and DNA/protein analysis.
Features
- Accurate analysis via a holographic grating that greatly reduces stray light.
- Powerful functionality including photometric measurement, quantitative measurement, spectrum scan, DNA/protein analysis, and printing capability, plus PC-integrated 3D spectrum analysis and GLP laboratory protocol support.
- Convenient operation — a high degree of automation means the operator only needs to press keys twice.
- Easily upgraded with optional accessories that expand flexibility and measurement range.
- Easy routine maintenance thanks to a simple mechanical structure and modular electrical design.
- Deuterium and tungsten lamps that are easily replaceable and pre-aligned, with a motorized automatic 8-cell holder supplied as standard.
Technical Specifications
| Specification | Details |
|---|---|
| Optics | Double Beam with Automatic 8 Cell Charger |
| Wavelength Range | 190-1100 nm |
| Spectral Band Width | 0.5, 1, 2, 5 |
| Wavelength Display | 0.01 nm |
| Wavelength Setting | 0.05 nm |
| Wavelength Accuracy | ± 0.1 nm @656.0 nm D2, ± 0.3 nm (190-1100) |
| Wavelength Repeatability | 0.1 nm |
| Stray Light | 0.03% @340 & 220 nm, 1% @198 nm for KCI |
| Photometric Range | -3 to 3 A |
| Photometric Accuracy | ±0.002 A at 0.5, ±0.004 at 1 A |
| Photometric Reproducibility | 0.001 A (0.5 A), 0.002 A (1A) |
| Baseline Stability | ±0.0002 Abs |
| Baseline Flatness | ±0.001 Abs |
| Scan Speed | 5-3500 nm/min |
| Slew Speed | 5000 nm/min |
Configuration Options
- Operates stand-alone or with a PC and accompanying software; software is supplied with the instrument.
Optional Accessories
- PS181-2 Sipper Pump Accessory (Pump, Tubing, Cassette, Front Panel, Flow Cell).
- CH188-1 Position 10mm Constant Temperature Cell Changer (for use with PTC-2).
- PTC-2 Peltier Module.
Note: We are Manufacturers, Suppliers, Exporters and Dealers of Labindia UV-3000 Spectrophotometer in Mumbai, Hyderabad, Bangalore, Pune, Ahmedabad, Chennai, Visakhapatnam, Surat, Nashik, Warangal, Vijayawada, Coimbatore, Vadodara, Navi Mumbai, Thane, Maharashtra, Goa, Gujarat, etc.